68
Other Procedures
15 Check the difference between the
results and decide whether correcting
the reference value of the sensor is
required.
If the difference is 0.3mm or less, no
adjustment is necessary.
If the difference exceeds the 0.3mm
permissible range, adjustment is
necessary.