6. BASIC OPERATION
32
Direct entry of grade value
(Example) Setting grade of -12.345% (Format ± AB.CDE%)
• Input range: -15.000 to 40.000% (-150.00 to 400.00‰)
• Setting grade is not possible when the instrument is locked.
6.2 Grade setting procedure
Procedure Key operation Display
1 Press [SET] key.
The previous data will be shown, and ±
mark is highlighted.
2 Press [ ] or [ ] key to change the sign
to -.