Why are the grade setting and measured grade value not equal on my Topcon Measuring Instruments?
- LLaura WallaceAug 13, 2025
There are several reasons why the grade setting value of the Topcon TP-L6W Measuring Instrument and the measured grade value might not match: * An incorrect grade value might have been entered. To resolve this, check the input value (‰ or %) and reset it. * The bubble on the bubble tube display of the instrument or the bubble on the target may not have been adjusted correctly. Adjust the laser and/or target so the bubble is centered. * The laser beam could be refracted due to temperature differences within the pipe. When laying the pipe, backfill the trench immediately to help maintain temperature stability. Consider using a blower to move air through the pipe to keep temperature and moisture stable.