8. CHANGING OPERATING PARAMETERS
52
ON/OFF of the
centerline LED/
upper and lower
lasers' automatic
cut-off (automatic
power off)
V-LED/
V-LD (GV/AV)
--- : Automatic cut-off is disabled. LED/LD
remains on whenever after being
turned on by user.
30 : Default setting. Centerline LED/LD
automatically cuts off 30 minutes after
it has been turned on by user.
p.19
Switch the unit
of the grade
values displayed
(% /
)
UNIT
% : Default setting. Grade is displayed to
the nearest thousandths of a percent
(+01.235%)
‰ : Grade is displayed as per mill
[+01.235% (percent) equals +012.35
(per mill)]
ON/OFF of
Security Mode
S CODE Activates security code function. Default is
OFF. Changing to ON requires the entry of a
4-digit security to turn on instrument.
Input / Change
the Security Code
CHANGE
S CODE
(no preset code)
Allows user to input any 4-digit security
code after S CODE is set to ON.
Change the
Company Name
CHANGE NAME
(TOPCON preset)
Allows user to enter name that will display
during powerup. Default name is TOPCON.
--- / [30]
% /
ON / [OFF]