EasyManua.ls Logo

Toshiba 4400 Series - Setup: Calibrating the Touchscreen

Toshiba 4400 Series
138 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
89
4400 Series Installation and Operation Manual – 64527-008
12.17 Setup: Calibrating the Touchscreen
At anytime while using the tabs to view system details, the touchscreen can be calibrated by executing the built-in
calibration routine.
To calibrate the touchscreen:
1. Press the Setup Tab.
2. Press the DISP-EX 800 button.
3. Press the entry Cali (Parameter 8211)
to begin the calibration routine.
4. Follow the directions on the display to
press the “+” at each location it ashes
on the display. To exit the calibration
routine press QUIT at any time.
See Fig. 12-15 A, B.
5. After the touchscreen has been recalibrated
the display will indicate that the calibration
routine has passed. Press Quit to return
to the main display. See Fig. 12-15 C.
FIGURE 12.15: TOUCHSCREEN DIS-
PLAY CALIBRATION
A
B
C

Table of Contents

Other manuals for Toshiba 4400 Series

Related product manuals