EasyManuals Logo

Tracerco NORM Monitor IS User Manual

Tracerco NORM Monitor IS
55 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #22 background imageLoading...
Page #22 background image
22 MN1007-C
4.3 Measurement Modes for the Scintillator Probe.
Following initialisation – see previous section - the user may choose to change the measurement unit mode. The
various measurement modes for the scintillator probe are:
cps count rate (or cpm)
background subtracted cps count rate (or cpm)
Dose rate in µSv/h (or µR/h)
background subtracted dose rate in µSv/h (or µR/h)
In each case it is possible to utilise the integration function whereby the displayed reading is averaged over a
period of time.
Note: The NORM Monitor-IS is capable of displaying SI units as well as non-IS units (primarily US market). Cps or
cpm & Sv/h or µR/h are factory set parameters and not configurable in the field.
4.3.1 cps Mode
Press the mode
key until the “cps text appears on the LCD display as shown.
Figure 5 - cps readout screen
This screen provides a bar graph and digital display indication of the radioactivity detected by the probe in counts
per second (cps).
In cps mode the value displayed indicates the total count rate detected by the probe and includes any contribution
from environmental background radioactivity.
The green indicator LED will flash whilst the instrument is counting in this measuring mode

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tracerco NORM Monitor IS and is the answer not in the manual?

Tracerco NORM Monitor IS Specifications

General IconGeneral
BrandTracerco
ModelNORM Monitor IS
CategoryMonitor
LanguageEnglish