43 MN1007-C
11 Essential Safety Information For Hazardous Areas
The NORM Monitor-IS handset and probes are safe for use in hazardous areas described in the equipment
marking and according to the essential safety information stated below.
Note: Approval dependant on product version ordered and described on equipment label.
Marking details
ATEX Certificate number: Baseefa12ATEX0209X
IECEx Certificate number: IECExBAS12.0114X
CSA Certificate number: 13.2616487
ATEX Equipment Marking: Ex ia IIC T4 Ga (-20°C Ta +50°C)
Equipment category: II 1G
CSA Equipment Marking:
Class I, Div 1, Groups A, B, C, D; Temp code T4
Class I, Zone 0, Ex ia, IIC, T4
Class I, Zone 0, AEx ia, IIC, T4
Approval standards
BS EN 60079-0:2012 – Explosive atmospheres. Equipment. General requirements
BS EN 60079-11:2012 – Explosive atmospheres. Equipment protection by intrinsic safety “i”
CSA C22.2; CAN / CSA / UL 60079-0; CAN / CSA / UL 60079-11; UL 913 7
th
Edition
Limitations of hazardous area use
ATEX coding II 1G states that the equipment is for non-mining use (II) with highest protection (1) for gas (G)
atmospheres.
Electrical Coding Ex ia IIC T4 Ga (-20°C Ta +50°C) states that the equipment is intrinsically safe (ia) for use in
zones 0, 1 and 2 and that it may be used in areas where flammable gas groups IIA, IIB, and IIC are present. T4
indicates that the equipment is suitable for use in flammable gasses with a T Class of T1, T2, T3 or T4. The
ambient working temperature for the equipment is -20°C to +50°C.
Condition of safe use
Transportation, removal and replacement of the scintillator from the PM probe (model SA-50/1) must be done in a
non-hazardous area only.
Practical safety points
Do not open the battery compartment in a hazardous area.
Use only the Duracell MX1604 or MN1604 alkaline manganese battery.
Lock the battery compartment by the screw in the compartment lid -do not over-tighten.
Clean with damp cloth. Do not use solvents on the handset monitor window.
Check the monitor and probes for signs of damage or cracks in the casing before use.
Manufacturer details:
Name and address of the manufacturer is printed in the product documentation as well as moulded into rear of the
monitor handset.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tracerco NORM Monitor IS and is the answer not in the manual?

Tracerco NORM Monitor IS Specifications

General IconGeneral
BrandTracerco
ModelNORM Monitor IS
CategoryMonitor
LanguageEnglish