8. PERPENDICULARITY MEASUREMENTS
Before taking any perpendicularity measurements, the
carriage must be blocked.
8.1 With Test Indicator
1. Place a test indicator in the probe holder.
2. Lightly press the lever of the indicator against the side
to be measured and set the zero.
3. Move the instrument's carriage vertically to check the
perpendicularity of the side.
To get optimum measuring precision, the instrument base must be completely
parallel to the side to be measured.
8.2 With Electronic Probe
The perpendicularity measurement is given with the
help of 4 values represented in the diagram alongside:
Distance
Perpendicularity
Inclination
Rectitude
V7
V9
Z
X