Using the numeric and decimal keys, enter the measured
density (the density as measured by the gauge) at the first test
location. Press the
〈ENTER/START〉 key. The gauge displays:
Enter the actual density (the density as determined from an
alternate method) and press the
〈ENTER/START〉 key. The
gauge displays:
Press the 〈STORE〉 key to store the values and to enter those
for the next location. When the measured and actual densities
for each location have been entered, press the
〈ENTER/START〉 key to calculate the mix calibration. The
gauge displays:
To return to the Setup menu without saving the mix
calibration, press the
〈NO〉 key.
To save the mix calibration, press the 〈YES〉 key. The gauge
requests a mix calibration ID as shown:
0. 0 pcf
Pr ess <ENTER>
<STORE> f or Nex t
<ENTER> t o Cal cul at e
I nt er cept = ###. #
Do you want t o
accept t hi s mi x?
<ALPHA> f or Let t er s
Pr ess <ENTER>