2 Getting Started
Up and Running Overview
Getting Started
Up and Running Overview
1.
Place the QUESTemp
º
48N in the work area in a safe location approximately 3.5 feet
off the ground.
2.
Turn the unit On. If the battery voltage displayed during the power-on sequence is less
than or equal to 6.4 volts, replace or recharge the batteries.
3.
Be aware that the sensors require 10 minutes to stabilize to a new environment.
4.
In the main menu, View will be selected (an indicator arrow denotes the selected
menu). Press the I/O Enter key and the measurement screen will appear.
5.
Press the Run/Stop key to begin datalogging. Use the Arrow keys to set the display to
the desired items.
Placing the QUESTempº 48N on the Job Site
The QUESTempº 48N should be placed at a height of 3.5 feet (1.1 m) for standing individuals
or 2 feet (.6 m) for seated individuals. Tripod mounting is recommended to get the unit away
from anything that might block radiant heat or airflow. A 1/4" x 20 threaded bushing on the
bottom of the instrument allows mounting to a standard photographic tripod. DO NOT stand
close to the unit during sampling.
Before data logging, allow ten minutes for the sensors readings to stabilize.
Getting Started