Operation and Service Manual 25
Calibration Offset Adjustments Applicable Models
960, 962,
964, 966,
980, 982,
800220
9600, 9630, 9650,
9650-NB
964, 966, 980, 982, 995,
800220
You can apply offset adjustments
to temperature, relative humidity,
and barometric pressure. The
measurement displayed and
logged will be the raw
measurement added to the value
you specify here. Offsets will be
shown on the dashboard if the
value is other than 1. Enter 0 to
return to default.
Correction Factors Applicable Models
You can apply correction factors to
pitot velocity, pitot flow and thermal
anemometer and rotating vane air
flows. The measurement displayed
and logged will be the raw
measurement multiplied by the
value you specify here. Correction
factors will be shown on the
dashboard if the value is other than
1. Enter 1.000 to return to default.
IAQ Probe Calibration Applicable Models