48 VelociCalc
®
/VeloCalc
®
Pro Air Velocity Meter 9600 Series
962 and 966 Thermoanemometer Probe Dimensions (962, 966)
Probe Length: ...............................
40 in. (101.6 cm)
Probe Diameter of Tip: .................
0.28 in. (7.0 mm)
Probe Diameter of Base: ..............
0.51 in. (13.0 mm)
Articulating Section Length: ..........
6.0 in. (15.24 cm)
Diameter of Articulating Knuckle: .
0.38 in. (9.5 mm)
Thermoanemometer Response Time
Velocity: ...................
200 msec
Temperature: ...........
2 minutes (to 66% of final value)
Humidity: ..................
<1 minute (to 66% of final value)
995 Rotating Vane 4 in. (100mm) Probe Velocity and
Temperature
Range: .....................
50 to 6000 ft/min (0.25 to 30 m/s),
32 to 140°F (0 to 60°C)
Accuracy: .................
±1% + 4 ft/min (±0.02 m/s), ±2.0°F (±1.0°C)
Resolution: ...............
1 ft/min (0.01 m/s), 0.1°F (0.1°C)
980 IAQ Probe CO
2
, Temperature and Humidity
Range: .....................
0 to 5000 ppm CO
2
, 5 to 95% RH,
14 to 140°F (-10 to 60°C)
Accuracy: .................
±3% of reading or ±50 ppm CO
2
, whichever
is greater
9
, ±3% RH
6
, ±1.0°F (±0.5°C)
5
Resolution: ...............
1 ppm CO
2
, 0.1% RH, 0.1°F (0.1°C)
982 IAQ Probe CO, CO
2
, Temperature and Humidity
Range: .....................
0 to 500 ppm CO, 0 to 5000 ppm CO
2
,
5 to 95% RH, 14 to 140°F (-10 to 60°C)
Accuracy: .................
±3% of reading or ±3 ppm CO, whichever is
greater
8
, ±3% of reading or ±50 ppm CO
2
,
whichever is greater
9
, ±3% RH
6
, ±1.0°F
(±0.5°C)
5
Resolution: ...............
0.1 ppm CO, 1 ppm CO
2
, 0.1% RH, 0.1°F
(0.1°C)
980 and 982 Probe Response Time
Carbon Dioxide: .......
20 seconds to 90% of final value.
Carbon Monoxide: ...
<60 seconds to 90% of final value.
Temperature: ...........
30 seconds (90% of final value, air velocity
at 400 ft/min [2 m/s])
Relative Humidity: ....
20 seconds (for 63% of final value)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments VelociCalc 9600 Series and is the answer not in the manual?

TSI Instruments VelociCalc 9600 Series Specifications

General IconGeneral
BrandTSI Instruments
ModelVelociCalc 9600 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals