50 VelociCalc
®
/VeloCalc
®
Pro Air Velocity Meter 9600 Series
Instrument Operating Conditions
Altitude up to 4000 meters (only limited when plugged into the
AC/DC adapter)
Relative humidity up to 80% RH, non-condensing
Pollution Degree II
Overvoltage Category II
Data Storage Capabilities
Range: ............................
162,200 samples and 200 TESTIDs
(one sample can contain multiple
measurement types)
Logging Interval
Intervals: .........................
1 second to 1 hour
Time Constant
Intervals: .........................
User selectable
External Meter Dimensions
3.2 in. 9.5 in. 1.6 in. (8.1 cm 24.1 cm 4.1 cm)
Meter Weight
Weight with Batteries: .....
0.9 lbs (0.41kg)
Power Requirements
Four (4) AA-size batteries (included) or AC/DC power supply
P/N 6013125
Input: ...............................
100 to 240 VAC, 50 to 60 Hz, 1.0A
Output: ............................
12 VDC, 3.0A
1
Temperature compensated over an air temperature range of 40 to 150°F
(5 to 65°C).
2
The accuracy statement of ±3.0% of reading or ±3 ft/min (±0.015 m/s), whichever is
greater, begins at 30 ft/min through 9999 ft/min (0.15 m/s through 50 m/s).
3
Pressure velocity measurements are not recommended below 1000 ft/min (5 m/s) and
are best suited to velocities over 2000 ft/min (10.00 m/s). Range can vary depending
on barometric pressure.
4
Accuracy is a function of converting pressure to velocity. Conversion accuracy
improves when actual pressure values increase.
5
Accuracy with instrument case at 77°F (25°C), add uncertainty of 0.05°F/°F
(0.03°C/°C) for change in instrument temperature.
6
Accuracy with probe at 77°F (25°C). Add uncertainty of 0.1% RH/°F (0.2% RH/°C) for
change in probe temperature. Includes 1% hysteresis.
7
Overpressure range = 190 in. H
2
O (7 psi, 360 mmHg, 48 kPa).
8
At 77°F (25°C). Add uncertainty of ±0.2%/°F (0.36%/°C) for change in temperature.
9
At calibration temperature. Add uncertainty of ±0.28%/°F (0.5%/°C) for change
in temperature.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the TSI Instruments VelociCalc 9600 Series and is the answer not in the manual?

TSI Instruments VelociCalc 9600 Series Specifications

General IconGeneral
BrandTSI Instruments
ModelVelociCalc 9600 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals