Questions and Answers

  • H
    hernandezgabrielAug 25, 2025
    What does ‘Low Heat’ mean on my Tuttnauer 2340EKA?
    • D
      Derek HollowayAug 26, 2025
      If the message ‘Low Heat’ is displayed during heat up on your Tuttnauer Laboratory Equipment, it indicates a bad temperature sensor. If heating elements are remaining on instead of cycling on and off, call for a technician.
  • K
    Kenneth GreenAug 26, 2025
    How to clean a very dirty chamber in my Tuttnauer 2340EKA?
    • K
      khopkinsAug 27, 2025
      If the chamber of your Tuttnauer Laboratory Equipment is very dirty, use Chamber Brite ™ to clean the autoclave; very dirty machines may require two cleanings.
  • J
    Jason BrownAug 27, 2025
    What to do if Tuttnauer 2340EKA Laboratory Equipment process fails during heating stage with a HIGH TEMP message?
    • J
      joshua59Aug 27, 2025
      If your Tuttnauer Laboratory Equipment process fails while initiating the heating stage and displays a HIGH TEMP message, try changing the setting of the thermostat.
  • C
    Carrie DayAug 27, 2025
    How to troubleshoot Tuttnauer 2340EKA machine not responding?
    • P
      Patrick ZimmermanAug 28, 2025
      If your Tuttnauer Laboratory Equipment machine isn't responding, verify that the main switch is in the ‘On’ position, the power cord is properly connected to both the machine and the mains, and the circuit breaker hasn't tripped. Also, check the reset button on the cut-out thermostat.
  • R
    Robert BatesAug 28, 2025
    What should I do if my Tuttnauer Laboratory Equipment shows ‘Low Pres’?
    • C
      Connor DeanAug 28, 2025
      If the 'Low Pres' message is displayed on your Tuttnauer Laboratory Equipment, clean the water level electrode inside the vessel. Also, ensure the door is fully closed, the door gasket is properly seated, and there is no steam leakage. Check the leveling of the machine.
  • J
    Jordan GarciaAug 29, 2025
    How to fix ‘High Pres’ error on Tuttnauer Laboratory Equipment?
    • E
      Edward BoydAug 29, 2025
      If the 'High Pres' message appears on your Tuttnauer Laboratory Equipment, the air jet might be clogged; clean it. It could also mean that the heating elements are continuously on, or the temperature sensor is faulty. In these cases, contact a technician.
  • J
    Jennifer McdonaldAug 30, 2025
    How to unclog the drain on Tuttnauer Laboratory Equipment?
    • B
      butlermichaelAug 30, 2025
      If the drain of your Tuttnauer Laboratory Equipment is clogged: 1. Open the drain by turning it counterclockwise and place a heavy object over the reservoir cover. 2. Blow compressed air into the drain to force debris back into the reservoir, then clean out the reservoir. 3. Alternatively, disassemble the drain valve, clean it, and blow out the line. Reassemble the valve, ensuring the large 'O' ring is properly seated.
  • C
    Christopher EwingAug 23, 2025
    What should I do if my Tuttnauer Laboratory Equipment displays ‘High Pres’?
    • J
      Joseph BergAug 23, 2025
      If your Tuttnauer Laboratory Equipment displays ‘High Pres’, clean the air jet. Ensure the sterilization time and temperature are set correctly; if unsure, use the default settings. Also, make sure the autoclave is not too heavily loaded.
  • M
    Marc MaysSep 1, 2025
    What to do if my Tuttnauer 2340EKA is not responding?
    • J
      Joel MossSep 1, 2025
      If your Tuttnauer Laboratory Equipment is not responding, ensure the main switch is in the ‘On’ position. Also, verify the power cord is properly connected to both the machine and the mains. Check the reset button on the cut-out thermostat and make sure the circuit breaker hasn't tripped; lift the circuit breaker lever if it has.
  • J
    Joseph ThompsonSep 2, 2025
    Why is my Tuttnauer 2340EKA leaking at the door?
    • K
      Kurt GarrettSep 2, 2025
      If your Tuttnauer Laboratory Equipment is leaking at the door, ensure the door is tightened enough and the door gasket is sealing the chamber. If necessary, replace the door gasket.

Summary

TECHNICAL DATA

Electrical Specifications

Lists electrical data, including voltage, amperage, and fuse ratings.

Technical Specifications

Lists detailed chamber dimensions, volumes, and pressure ratings.

KEYBOARD AND DISPLAY

Control Panel Key Functions

Describes the operation and functions of the control panel keys.

Displayed Messages and Safety

Lists error messages, their causes, and associated safety measures.

STERILIZATION PROGRAMS

Unwrapped Instruments Program

Details sterilization parameters and sequence for unwrapped items.

INSTALLATION AND SETUP

Electrical and Unit Setup

Covers electrical connection, safety, and initial unit configuration.

OPERATING INSTRUCTIONS

MAINTENANCE INSTRUCTIONS

Preventive Maintenance Schedule

Outlines daily, weekly, and periodic maintenance tasks.

TROUBLESHOOTING GUIDE

Initial Machine and Environment Checks

Covers machine response, heat, water level, and pressure issues.

Cycle, Seal, and Straining Problems

Addresses issues with exhaust time, door seals, clogged strainers, and drain problems.

Sterilization Process and Item Issues

Solutions for failing spore tests, wet/burned wraps, instrument rusting, and drying problems.