OPERATION
&
MAINTENANCE
MANUAL
Electronic Tabletop Autoclaves
Models 1730, 2340, 2540, 3140, 3850, 3870
E, EK, EA & EKA

Other manuals for Tuttnauer 2340EKA

Questions and Answers

  • A
    Andrew AlvaradoSep 2, 2025
    How to fix a Tuttnauer 2340EKA chamber that is not filling?
    • J
      James BrownSep 3, 2025
      If the chamber of your Tuttnauer Laboratory Equipment is not filling, check the water level in the reservoir. Also, inspect the bottom of the reservoir for any obstruction blocking the pickup tube. Clean the water sensor and try filling the machine manually using the Manual Inlet key.
  • M
    Michael DiazSep 3, 2025
    Why is the drying cycle taking so long in my Tuttnauer Laboratory Equipment?
    • A
      Aaron BrownSep 4, 2025
      If the drying cycle in your Tuttnauer Laboratory Equipment is taking too long, the autoclave may be overloaded. Also, consider replacing the HEPA filter.
  • F
    fowlerlisaSep 4, 2025
    Why does the exhaust stage take so long when running a cycle on my Tuttnauer 2340EKA?
    • R
      rileychristianSep 4, 2025
      If you are running a ‘glassware’ program on your Tuttnauer Laboratory Equipment, this is normal. The slow exhaust will take between 15 and 20 minutes.
  • G
    georgelewisSep 5, 2025
    Why is there a loud noise coming from the reservoir of my Tuttnauer 2340EKA?
    • M
      Mary KleinSep 5, 2025
      If there is a loud noise coming from the reservoir of your Tuttnauer Laboratory Equipment, the sterilization temperature might be set too high. The maximum setting is 274ºF (134ºC).
  • M
    Marcia BowenSep 6, 2025
    How do I fix a Tuttnauer 2340EKA printer that doesn't print?
    • B
      Benjamin CookSep 6, 2025
      If the printer of your Tuttnauer Laboratory Equipment does not print, ensure the paper is inserted in the printer. If the printer prints a test printout, the printer is O.K. and there is a problem with the electronics. Contact your dealer to solve the problem. If the printer does not print the test printout, there is a problem with the printer. Contact your dealer to solve the problem.

Summary

TECHNICAL DATA

Electrical Specifications

Lists electrical data, including voltage, amperage, and fuse ratings.

Technical Specifications

Lists detailed chamber dimensions, volumes, and pressure ratings.

KEYBOARD AND DISPLAY

Control Panel Key Functions

Describes the operation and functions of the control panel keys.

Displayed Messages and Safety

Lists error messages, their causes, and associated safety measures.

STERILIZATION PROGRAMS

Unwrapped Instruments Program

Details sterilization parameters and sequence for unwrapped items.

INSTALLATION AND SETUP

Electrical and Unit Setup

Covers electrical connection, safety, and initial unit configuration.

OPERATING INSTRUCTIONS

MAINTENANCE INSTRUCTIONS

Preventive Maintenance Schedule

Outlines daily, weekly, and periodic maintenance tasks.

TROUBLESHOOTING GUIDE

Initial Machine and Environment Checks

Covers machine response, heat, water level, and pressure issues.

Cycle, Seal, and Straining Problems

Addresses issues with exhaust time, door seals, clogged strainers, and drain problems.

Sterilization Process and Item Issues

Solutions for failing spore tests, wet/burned wraps, instrument rusting, and drying problems.