OPERATION
&
MAINTENANCE
MANUAL
Electronic Tabletop Autoclaves
Models 1730, 2340, 2540, 3140, 3850, 3870
E, EK, EA & EKA

Other manuals for Tuttnauer 3140E

Questions and Answers

  • J
    Julie HarrisAug 25, 2025
    What does ‘Low Heat’ mean on my Tuttnauer 3140E Laboratory Equipment?
    • J
      jacqueline62Aug 26, 2025
      If the message ‘Low Heat’ is displayed during heat up on your Tuttnauer Laboratory Equipment, it indicates a bad temperature sensor. If heating elements are remaining on instead of cycling on and off, call for a technician.
  • L
    lauren55Aug 26, 2025
    How to clean a very dirty chamber in my Tuttnauer 3140E?
    • V
      Virginia DoyleAug 26, 2025
      If the chamber of your Tuttnauer Laboratory Equipment is very dirty, use Chamber Brite ™ to clean the autoclave; very dirty machines may require two cleanings.
  • D
    Daisy HoldenAug 27, 2025
    What to do if Tuttnauer 3140E Laboratory Equipment process fails during heating stage with a HIGH TEMP message?
    • W
      wangmichaelAug 27, 2025
      If your Tuttnauer Laboratory Equipment process fails while initiating the heating stage and displays a HIGH TEMP message, try changing the setting of the thermostat.
  • W
    wbeckAug 27, 2025
    How to troubleshoot Tuttnauer Laboratory Equipment machine not responding?
    • O
      orichardsonAug 28, 2025
      If your Tuttnauer Laboratory Equipment machine isn't responding, verify that the main switch is in the ‘On’ position, the power cord is properly connected to both the machine and the mains, and the circuit breaker hasn't tripped. Also, check the reset button on the cut-out thermostat.
  • M
    Matthew OwensAug 28, 2025
    What should I do if my Tuttnauer Laboratory Equipment shows ‘Low Pres’?
    • C
      Christopher GreeneAug 29, 2025
      If the 'Low Pres' message is displayed on your Tuttnauer Laboratory Equipment, clean the water level electrode inside the vessel. Also, ensure the door is fully closed, the door gasket is properly seated, and there is no steam leakage. Check the leveling of the machine.
  • S
    Shannon GarciaAug 29, 2025
    How to fix ‘High Pres’ error on Tuttnauer Laboratory Equipment?
    • D
      Dustin FigueroaAug 29, 2025
      If the 'High Pres' message appears on your Tuttnauer Laboratory Equipment, the air jet might be clogged; clean it. It could also mean that the heating elements are continuously on, or the temperature sensor is faulty. In these cases, contact a technician.
  • J
    jessicajacksonAug 30, 2025
    How to unclog the drain on Tuttnauer Laboratory Equipment?
    • R
      reedcourtneyAug 30, 2025
      If the drain of your Tuttnauer Laboratory Equipment is clogged: 1. Open the drain by turning it counterclockwise and place a heavy object over the reservoir cover. 2. Blow compressed air into the drain to force debris back into the reservoir, then clean out the reservoir. 3. Alternatively, disassemble the drain valve, clean it, and blow out the line. Reassemble the valve, ensuring the large 'O' ring is properly seated.
  • M
    Matthew NelsonAug 23, 2025
    What should I do if my Tuttnauer 3140E Laboratory Equipment displays ‘High Pres’?
    • A
      Adam YoungAug 23, 2025
      If your Tuttnauer Laboratory Equipment displays ‘High Pres’, clean the air jet. Ensure the sterilization time and temperature are set correctly; if unsure, use the default settings. Also, make sure the autoclave is not too heavily loaded.
  • D
    dayalaSep 1, 2025
    What to do if my Tuttnauer 3140E Laboratory Equipment is not responding?
    • M
      Mary MullenSep 1, 2025
      If your Tuttnauer Laboratory Equipment is not responding, ensure the main switch is in the ‘On’ position. Also, verify the power cord is properly connected to both the machine and the mains. Check the reset button on the cut-out thermostat and make sure the circuit breaker hasn't tripped; lift the circuit breaker lever if it has.
  • H
    huntpaulSep 2, 2025
    Why is my Tuttnauer 3140E Laboratory Equipment leaking at the door?
    • A
      Alexander ScottSep 2, 2025
      If your Tuttnauer Laboratory Equipment is leaking at the door, ensure the door is tightened enough and the door gasket is sealing the chamber. If necessary, replace the door gasket.

Summary

General Information and Warranty

Autoclave Warranty Coverage

Details the manufacturer's warranty terms for components, labor, and chamber.

Warranty Activation and Statement

Procedures for activating the warranty and important statement of terms.

Technical Data and Specifications

Electrical Data and Fuse Information

Provides detailed electrical specifications, including circuit breakers and fuses.

Water Quality Requirements

Specifies acceptable physical characteristics and contaminant levels for autoclave water.

Control Panel and Display

Indicator Light Functions

Explains the meaning and function of various indicator lights on the control panel.

Control Panel Keyboard Functions

Details the program keys and other function keys on the control panel.

Display Panel Information

Description of the control panel screen layout and displayed information.

Displayed Messages and Safety Measures

Lists and explains system messages and associated safety measures for user guidance.

Sterilization Programs Guide

Program 1: Unwrapped Instruments

Details sterilization parameters and operational sequence for unwrapped items.

Program 2: Wrapped Instruments & Porous Loads

Details sterilization parameters and operational sequence for wrapped items and porous loads.

Program 3: Glassware Sterilization

Details sterilization parameters and operational sequence for glassware.

Installation Instructions

Electrical Connection and Setup

Guidance on electrical connections, safety, and initial setup for different models.

Automatic Water Filling System

Instructions for setting up and verifying the automatic water filling system.

Loading and Unloading the Device

Guidelines for safely loading and unloading materials into the autoclave chamber.

Filling the Water Reservoir

Instructions on correctly filling the autoclave's water reservoir with distilled water.

Preparation Before Sterilization

Instrument and Material Preparation

Guidelines for cleaning, drying, and packaging items for effective sterilization.

Operating Instructions

Maintenance Instructions

Preventive and Scheduled Maintenance

Outlines daily, weekly, and periodic maintenance tasks to ensure optimal performance.

Daily Maintenance Tasks

Specific daily cleaning and inspection procedures for the autoclave.

Weekly Maintenance Tasks

Specific weekly cleaning and descaling procedures for the autoclave.

Periodic Maintenance Tasks

Lists monthly and periodic checks, cleaning, and part replacements.

Troubleshooting Guide

Failed Sterilization Indicators

Solutions for when spore tests or indicator strips do not change color.

Tuttnauer 3140E Specifications

General IconGeneral
BrandTuttnauer
Model3140E
CategoryLaboratory Equipment
LanguageEnglish