Operation and
Maintenance Manual
Electronic Laboratory Autoclaves
Models 3840, 3850, 3870 ELVG-D
ELVG-D: standard autoclave
C: optional system for fast cooling
PV: optional vacuum pump
BH: optional bio hazard filtration
F: optional fan for super-fast cooling
Cat. No. MAN205-0448004EN Rev H
Tuttnauer Europe B.V.,
Hoeksteen 11 4815 PR P.O. Box 7191 4800 GD Breda,
The Netherlands Tel: 31 (0) 765423510, Fax: 31 (0) 765423540

Table of Contents

Other manuals for Tuttnauer 3850 ELVG-D

Questions and Answers

  • W
    wigginsadamAug 31, 2025
    What does 'Air Error' mean on Tuttnauer 3850 ELVG-D Laboratory Equipment?
    • A
      abrownAug 31, 2025
      If your Tuttnauer Laboratory Equipment displays 'Air Error' at the end of the cycle, it means the autoclave did not reach atmospheric pressure after 10 minutes. Wait until the autoclave reaches atmospheric pressure and then perform a new cycle.
  • C
    Carlos AllenSep 9, 2025
    How to fix Tuttnauer 3850 ELVG-D printer that doesn't print anything on the paper?
    • R
      riddlecharlesSep 10, 2025
      If your Tuttnauer Laboratory Equipment printer prints, but nothing appears on the paper, it likely means the paper roll isn't installed correctly. Ensure you install the paper roll the right way, noting that only one side of the paper is printable.
  • S
    Sherry CervantesAug 30, 2025
    What does 'High Pressure (Ending)' mean on my Tuttnauer Laboratory Equipment?
    • J
      Jimmy JordanAug 30, 2025
      If your Tuttnauer Laboratory Equipment displays 'High Pressure (Ending)', it means the system couldn't reach atmospheric pressure ± 10kPa during the ending stage. Perform a new cycle.
  • W
    Wendy PowellSep 13, 2025
    How to fix slow exhaust phases on Tuttnauer Laboratory Equipment?
    • D
      Daniel JosephSep 13, 2025
      If the fast or slow exhaust phases aren't functioning correctly, leading to high chamber pressure or a prolonged exhaust time, consider these potential causes: * The fast or slow exhaust valves might be stuck or damaged. Repair or replace any damaged valves. * The steam valve could be leaking. Repair or replace the steam valve. * There might be a control system issue. Replace the control boards.
  • R
    Ralph TorresSep 12, 2025
    Why is my Tuttnauer 3850 ELVG-D Laboratory Equipment leaking at the door?
    • K
      Kristen WarnerSep 12, 2025
      If your Tuttnauer Laboratory Equipment is leaking at the door, it could be due to: 1. A dirty door gasket. Clean the door gasket. 2. A damaged door gasket. Call for service.
  • S
    Stacy BrownSep 16, 2025
    Why are my Tuttnauer Laboratory Equipment sterilization results unsatisfactory?
    • A
      Austin NunezSep 16, 2025
      If you're experiencing unsatisfactory sterilization results, consider the following possible causes: * There might be incomplete removal of air pockets. * Sterilization conditions might involve improper temperature or pressure being too low, or the sterilization time being too short. Set the correct temperature and time according to the type and quantity of material. * Incorrect loading of the autoclave may be trapping air in the load.
  • A
    amymillerSep 23, 2025
    Why is the air detector failing in my Tuttnauer Laboratory Equipment?
    • V
      Victoria JarvisSep 23, 2025
      If the air detector fails, consider the following possible causes: * The steam quality might not be proper. Check the steam quality and non-condensable gases. * The induced air valve might be open or leaking. Close or repair the induced air valve.
  • J
    Jose MillerAug 27, 2025
    What does 'Low Pressure' mean on my Tuttnauer 3850 ELVG-D Laboratory Equipment?
    • K
      Kathleen RoseAug 27, 2025
      If your Tuttnauer Laboratory Equipment displays 'Low Pressure', it means the chamber pressure dropped below the sterilization pressure (134°C = 304 kPa, 121°C = 205 kPa) for 2 seconds during the sterilization stage. Try performing a new cycle.
  • S
    Seth SanchezAug 26, 2025
    What does 'High Temp' mean on my Tuttnauer 3850 ELVG-D?
    • D
      Douglas BrownAug 26, 2025
      If the 'High Temp' message is displayed on your Tuttnauer Laboratory Equipment, it means the temperature rose 7°F (4°C) above the sterilization temperature during the sterilization stage for 2 seconds during the sterilization cycle. Perform a new cycle.
  • M
    Michelle McdowellAug 25, 2025
    What does 'Low Temp' mean on my Tuttnauer 3850 ELVG-D Laboratory Equipment?
    • J
      Justin ArmstrongAug 26, 2025
      If your Tuttnauer Laboratory Equipment displays 'Low Temp', it indicates that the temperature dropped below the sterilization temperature for more than 1 second during the sterilization cycle. Try performing a new cycle.

Summary

Safety Instructions for Autoclave Operation

Technical Data and Specifications

Control Panel and User Interface

Display Screens and Messages

Information on display screens, operational messages, and error displays.

Screen Displays During Operation

Aborted and Failed Cycle Screens

Displays for aborted, canceled, or failed sterilization cycles.

Operation and Parameter Management

Parameter and System Settings

Changing operational parameters, system settings, and maintenance procedures.

Preparation and Loading Procedures

Autoclave Operating Instructions

Cycle Selection, Starting, and Error Handling

How to select programs, start cycles, and handle errors.

Unloading and Device Handling

Safe procedures for unloading the autoclave and moving the device.

Service and Maintenance Instructions

Scheduled Maintenance Tasks

Preventive maintenance for daily, weekly, and periodic checks.

Troubleshooting Common Issues

Machine, Printer, and Door Issues

Solutions for machine not responding, printer malfunctions, and door leaks.

Program, Supply, and Sensor Errors

Troubleshooting for vacuum, temperature, pressure, supply, and sensor errors.

Tuttnauer 3850 ELVG-D Specifications

General IconGeneral
TypeAutoclave
Chamber Volume85 L
Power Supply220-240V, 50/60Hz
Chamber MaterialStainless Steel
Temperature Range105°C - 137°C
Safety FeaturesSafety valve, Door protection interlock, Overheat protection