35
Corrections
4.1 Have a technician replace the “Dry” light.
5.1 Turn the multi-purpose valve to the DRY
position.
5.2 Turn the timer on.
5.3 Have a technician check out the unit.
6.1 Fill the reservoir with distilled water.
6.2 Follow the instructions in para 8.6 to unclog
the valve.
6.3 Follow the instructions in para 8.6 to unclog
the piping.
7.1 Clean the air jet see para 8.3
7.2 Adjust loading according to "Maximum
Load Sizes" table in sec. 2.
7.3 Items to be sterilized should be separated
and not stacked. See preparations for
sterilization para. 6. A Tuttnauer Pouch
Rack is recommended
7.4 Check table para 7.
7.5 See para. 10.9 “unit does not reach the
proper operating pressure”.
Possible cause check-up and tests
4.1 The “Dry” light is burned out.
5.1 The multi-purpose valve is not in the DRY
position.
5.2 The timer is off.
5.3 There is an internal problem.
6.1 There is no water in the reservoir.
6.2 The multi-purpose valve is clogged.
6.3 The fill piping is clogged.
7.1 Air jet is clogged.
7.2 Maybe over loaded.
7.3 The load may be to densely packed.
7.4 Sterilization time may not be correct.
7.5 Unit not reaching the correct sterilization
pressure reading on the gauge.
Symptom
4. Dry indicator light
does not light up at
the beginning of the
dry cycle. The
Power light is on and
the unit does heat
up.
5. Dry indicator light
does not light up at
the beginning of the
dry cycle. The
Power light is on and
the unit does NOT
heat up.
6. Water does not enter
the Chamber when
the multi-purpose
valve is in the FILL
position.
7. Spore tests are
failing or indicator
strips are not
changing color
according to
instructions of
indicator's
manufacturer.