EN 21
75kg
1017 x 825 x 1225
IP 20
Table 3: Technical data
OMEGO
®
Basic is classified as a medical electronic device and therefore
subject to specific precautionary measures relating to electromagnetic
compatibility (EMC). It is absolutely imperative to observe the stated EMC
indications. Portable and mobile HF communication equipment may affect
OMEGO
®
Basic.
Guidelines and MANUFACTURER‘s declaration – ELECTROMAGNETIC
EMISSIONS
The OMEGO® Basic system is designed for operation in an ELECTROMAGNETIC ENVIRONMENT as indicated
below. The customer or user of the OMEGO® Basic system must ensure that it is used in such an
environment.
Interference emission
measurements
Agreement
ELECTROMAGNETIC
ENVIRONMENT
Guidelines
HF emissions according to
CISPR 11
Group 1
The OMEGO® Basic system
exclusively uses HF energy for its
internal FUNCTIONS. HF emissions
are very low and unlikely to disrupt
electronic devices within range.
HF emissions
according to CISPR 11
Class B
The OMEGO® Basic system is
suitable for usage in all
establishments including
residential areas and areas that
are directly connected to the
PUBLIC SUPPLY GRID, which also
supplies residential buildings.
Harmonics emissions according
to IEC 61000-3-2
Class A
Emissions of voltage
fluctuations/flicker according to
IEC 61000-3-3
Not applicable
Tabelle 4: Guidelines and manufacturer‘s declaration Electromagnetic emissions