Table 5: Guidelines and Manufacturers‘s declaration Electromagnetic Immunity
Guidance and Manufacturer’s Declaration ELECTROMAGNETIC
IMMUNITY
The OMEGO® Basic system is designed for operation in an ELECTROMAGNETIC ENVIRONMENT as
indicated below. The customer or user of the OMEGO® Basic system must ensure that it is used in such an
environment.
Portable and mobile RF communications equipment should be used no closer to any part of the OMEGO®
Basic system, including cables, than the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Immunity Test
IEC 60601-Test
Level
Compliance Level
ELECTROMAGNETIC
ENVIRONMENT - Guidance
Recommended Separation
Distance:
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 V
𝑑 = 1,17
𝑃
Radiated RF
IEC 61000-4-3
10 V/m
80 MHz to 2,7 GHz
10 V/m
𝑑 = 0,35
𝑃 80 MHz to 800
MHz
𝑑 = 0,7
𝑃 800 MHz to 2,5
GHz
where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site
survey
a
, should be less than the
compliance level in each
frequency range.
b
Interference may occur in the
vicinity of equipment marked
with the following symbol: