Manual CWD 2005 Software
___________________________________________________________________
Manual CWD2005.doc 47
6.2.1.1.3. Display mA
This menu displays the mA output of each of the installed output modules in the
information display. In the screen example, the first output is selected (using the
an be installed. This
feature can be helpful for output signal loop tests to remote equipment.
.
6.2.1.1.4. Digital inputs
Signal and zero position are multiple choice list fields. These inputs can start a
calibration cycle, or switch on and off an output hold module. The inputs are polled
once a second and any change of status must apply for at least one second to be
recognized by the system.
Signal
Zero position
Contact
Start
measurement
high
open
closed
low
open
closed
Start calibration
high
open
closed
low
open
closed
Hold signal
high
open
closed
low
open
closed