6.2.1.3.2. Ignition

same as the ignition duration. The longest ignition duration is 100 seconds. Normally
speaking, an ignition cycle should last between 15 and 20 seconds.

degree count that the thermopile recognizes the flame is burning.
Flame temperatures are gas composition dependant so the factory will preset the
optimum temperature for the specified measuring range. In the example, 3°C is the
differential temperature of the incoming cooling air/flue gas mix at the thermopile.
6.2.1.3.3. Update
If a new update is required, a CWD 2005 software memory stick has to be inserted in
the left side             
starts automatically.
 memory stick
(copy time is about 1 minute).
The CWD 2005 boots automatically and starts the measurement operation.
Note: without
a memory stick in the USB port.