Software Manual CWD 2005
___________________________________________________________________
50 Manual CWD2005.doc
6.2.1.2.1. Configure Calibration gas
Calibration gas data is entered as a Wobbe Index and Specific Gravity number. If the
gas vendor has only supplied the component % then the user must calculate and
enter these three calibration gas values from the components of the calibration gas.
A small software program (excel file) can be obtained from the factory, which can
also be downloaded from the Internet homepage. Also the memory stick which is
delivered with the instrument obtains this program.This includes most of the gases
found in many common process gas streams and this utility will help speed and any
necessary calculations.
6.2.1.2.2. Automatic calibration