Circuit diagrams Manual CWD 2005
___________________________________________________________________
94 Manual CWD2005.doc
10.2.1.1. Relay outputs
The relay outputs are connected to terminal block X14.
The relay contacts have the following specifications
Maximum current: 1amp max. 100 Volt, 100V DC/30Watt, 100V AC/60VA
If the relay is operational, a red alarm LED is illuminated.
The functional position for the relay outputs can be user configured from the menu
and the standard factory configuration is shown below.
Block X14 on I/O Card EA External type_06
Position Terminals Function Alarm LED
1 1 COM Common Process D 2
2 NO norm.open
3 NC norm.closed
2 4 COM Maintenance D 4
5 NO
6 NC
3 7 COM Filter change D 5
8 NO
9 NC
4 10 COM Fault D 6
11 NO
12 NC
5 13 COM Wobbe I <XXXX D 7
14 NO
15 NC
6 16 COM Wobbe I <XXXX D 8
17 NO
18 NC
7 19 COM Customer-specific pos. D 9
20 NO
21 NC
8 22 COM Customer-specific pos. D 10
23 NO
24 NC