Troubleshooting Manual CWD 2005
___________________________________________________________________
82 Manual CWD2005.doc
Note: After several minutes of loading files the screen will return to the start menu
with the six vertical arrows, power down the unit, remove memory stick and restore
power. The unit will now run through the normal short start-up routine.
This is a recovery memory stick.
With this memory stick you can reboot or repair the system after a crash
You can reformat the "Disk on Chip" for regenerating the system.
You can install different updates and you can save any instrument specific data.
Each command name points with an arrow to a push button.
Caution:
The program starts automatically and after finishing the routine (ignore any other screen
prompts!) will return to the screen shown below, remove memory stick and turn the
Power switch off then on again. The software starts normally, with a basic configuration
or with individual configurations if the memory stick contains this data.
l l l l l I
Full update
This program should only be initiated after consulting the factory. This program
updates the complete software with standard factory data or the last stored data.
individual calibration data is loaded from the computer to the memory stick, so that
6 should always be executed, when the customer changes configuration parameters.