EasyManua.ls Logo

UP3D ACUBLU - Calibration; Step of Calibration

Default Icon
65 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
UP3D TECH CO.,LTD
12
1.5 Calibration
Calibration will affect accuracy. Calibration is generally required in the following situations:
The first use of scanner; or use again after long period of time.
After opening the calibration software for the first time or starting the scan through the dental
manager, the hint window (figure 1-6) will pop up. Click re-calibration.
2When starting calibration for the first time, if the software detects that the camera needs to
initialize, then a dialog box (figure 1-7) will pop up and click the "ok" button. The software will
automatically adjust the camera and enter the calibration interface again
When the scanner is used again after transportation.
When there is rough surface or abnormal data quality in the scanning process
Figure 1-6
Figure 1-7
1.5.1 Step of calibration
1Double-click the ‘UP3D Scan Calibration’ shortcut or click the Calibrate scanner button at
the dental manager interface to start the Calibration function.
2insert the calibration board and use the shading cloth. Click ‘start’ (figure 1-8) to enter the
calibration interface

Related product manuals