Autoclaving
The collection bottle with bottle head and screw
cap, the quick coupling and the lter are des-
ignated for steam sterilization at 121°C and 2
bar absolute (1 bar overpressure). Time of ex-
posure according to DIN 58946 t
e
= 20 minutes.
Prior to autoclaving loosen or remove the
bottle head from the bottle.
It is the user´s responsibility to ensure effective
autoclaving.
Cleaning and decontaminating
121°C
2 bar absolute
The number of autoclaving cycles can be
marked on the plastic disc (Memory Disc) of the
lter (max. 20 autoclaving cycles according to
the manufacturer instruction).
Autoclaving
Maximum liquid level in collection bottle:
80%. When working with low boiling liquids or
liquids that tend to foam, the maximum liquid
level may be less.
max.
80%
121°C
2 bar
absolute
max. 20 x
NOTICE
NOTICE
M
i
d
i
s
a
r
t
®
2
0
0
0
0
.
2
0
µ
m
P
T
F
E
IN
sartorius stedim
Autoclaving
Over time, discoloration and material changes (e.g., re-
siliency, elasticity / tightness, cracking) due to repeated
steam sterilizations / autoclavings and / or chemical disin-
fections may occur.
Check all parts regularly. Replace defective parts.

Table of Contents

Other manuals for vacuubrand BVC basic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the vacuubrand BVC basic and is the answer not in the manual?

vacuubrand BVC basic Specifications

General IconGeneral
Brandvacuubrand
ModelBVC basic
CategoryWater Pump
LanguageEnglish

Related product manuals