Chapter 4 ________________________________________________________________ Operation
VAISALA______________________________________________________________________ 105
Example:
>pres
Pressure : 1013.00 hPa ?
>pres 2000
Pressure : 2000.00 hPa
>
Table 27 Conversion Factors for Pressure Units
From To: hPa
mbar 1
Pa N/m2 0.01
mmHg torr 1.333224
inHg 33.86388
mmH
2
O 0.09806650
inH
2
O 2.490889
atm 1013.25
at 980.665
bar 1000
psia
1)
68.94757
1) psia = psi absolute.
For example, 29.9213 inHg = 29.9213 × 33.86388 = 1013.25 hPa.
Date and Time
Using Display/Keypad
If the optional Data Logger Module is installed, you can change the time
and date using the display/keypad.
1. Press any of the arrow buttons to open the Main Menu.
2. Select System and press the arrow button to confirm your
selection.
3. Select Date and time and press the arrow button.
4. Press the SET button to enter the adjustment mode, and use the
arrow buttons to select and change the values.
5. You can also change the date and time formats that are shown in
the graphs. The selected formats are only used in graphical display,
they do not change the formats that are used in the serial
communication.
6. Press EXIT to return to the basic display.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vaisala HUMICAP HMT330 SERIES and is the answer not in the manual?

Vaisala HUMICAP HMT330 SERIES Specifications

General IconGeneral
Humidity Measurement Range0 ... 100 %RH
Protection ClassIP65
Measurement ParametersRelative Humidity, Temperature
Temperature Measurement Range-40 to +60 °C
Humidity Accuracy±1 %RH (0 ... 90 %RH) ±1.7 %RH (90 ... 100 %RH)
Temperature Accuracy±0.2 °C
Humidity SensorHUMICAP®
Output SignalsAnalog
Power Supply10...35 VDC
Operating Temperature Range-40 to +60 °C