User's Guide ______________________________________________________________________
156 _________________________________________________________________ M210566EN-J
0508-013
Figure 77 Adjustment and Purge Buttons
The following numbers refer to Figure 77 above:
1 = Indicator
LED
2 = Adjustment button
3 = Press the purge buttons simultaneously to start chemical
purge (if available)
Adjustment menu is displayed only when ADJ button (on the
motherboard inside the transmitter) is pressed.
0706-007
Figure 78 Adjustment Menu
Table 33 Indicator Led Functions
Indicator Led Function Description
LED off adjustment locked
LED on adjustment available
LED blinking evenly measurement not stabilized
LED blinking with short pulses performing chemical purge
NOTE
If using a warmed probe (HMT337 option), probe heating will be
interrupted when ADJ key is pressed. Allow sufficient time for the
probe to reach ambient temperature before starting the adjustment
procedure.
NOTE
Fixed pressure compensation value of 1013.25 hPa is used when in
adjustment mode