Chapter 6 ________________________________________________ Maintenance
VAISALA__________________________________________________________47
0205-020
Figure 12 Removing the LMB201 and LM211246 Units
The following numbers refer to Figure 12 above:
1 = LMB201 mounting screws
2 = LMB201 Measurement Board
3 = Heater and window contamination connectors
4 = LM211246 Optics Holding Assembly
5 = LM211246 Mounting screws
6 = LM211244 Cover Frame Subassembly and Cabling
6. Take the new Cover Subassembly and Cabling unit
(LM211244) and mount the old Optics Holding
Assembly (LM211246) into it with the four screws.
7. Connect the heater and window contamination
connectors onto the LMB201 board and mount the
board in place onto the Optics Holding Assembly. The
sensitivity measurement LED has to be inserted into
the correct opening in the Optics Holding Assembly,
see Figure 13 on page 48. Secure the LMB201 board in
place with the three retaining screws.