12. Look at the display and wait for the data logger to start up. You should see the word NEW
on channel 2 for a few seconds, after which it is replaced by the temperature reading.
13. Attach a label with number 2 to the probe you just connected.
14. Look at the display:
If dashes "- - -" are shown instead of measurement readings, check that the probe and
probe splitter are properly connected. It is possible to disconnect the probe splitter
accidentally by pulling it with the battery cover open.
Battery indicator should show full batteries
.
There should be no error codes shown. If there are, see Error codes (page 57).
15. Close the battery cover of the data logger. Push the latch down until you hear a click. If
the cover does not close easily, push the probe splitter in and try again.
Once the data logger has been set up in two probe mode, it will require a factory
reset to allow it to be set up in one probe mode again.
More information