Property Specification
With mounting bracket 186 × 68 × 36.5 mm (7.32 × 2.68 × 1.44 in)
Weight
With batteries (2 pcs alkaline) and HMP115 probe 190 g (6.7 oz)
With batteries (2 pcs alkaline), HMP115 probe,
and magnetic mounting bracket
254 g (8.96 oz)
Materials
Housing PC/ABS blend
Display window PMMA (acrylic)
Sealings TPE
7.2 RFL100 compatible probes technical
specification
Table 18 HMP110/T probe measurement performance
Property Description/Value
Relative humidity
Measurement range 0 … 100 %RH
Accuracy in temperature range 0 … +40 °C (+32 … +104 °F)
1)
0 … 90 %RH
90 … 100 %RH
±1.5 %RH
±2.5 %RH
Accuracy in temperature range −40 … 0 °C, +40 … +80 °C (−40 … +32 °F, +104 … +176 °F)
1)
0 … 90 %RH
90 … 100 %RH
±3.0 %RH
±4.0 %RH
Factory calibration uncertainty at +20 °C (68 °F)
2)
0 … 90 %RH
90 … 100 %RH
±1.1 %RH
±1.8 %RH
Humidity sensor HUMICAPâ 180R
Stability ±2 %RH over 2 years
Temperature
Measurement range −40 … +80 °C (−40 °F … +176 °F)
Accuracy over temperature range
Chapter 7 – Technical data
63