Property Description/Value
at 0 … +40 °C (+32 °F … +104 °F) ± 0.2 °C (0.36 °F)
at −40 … 0 °C, +40 … +80 °C (−40 … +32 °F, +104
… +176 °F)
± 0.4 °C (0.72 °F)
Factory calibration uncertainty
2)
± 0.2 °C (0.36 °F)
Temperature sensor Pt1000 RTD Class F0.1 IEC 60751
1) Includes non-linearity, hysteresis, and repeatability.
2) Small variations possible; see also calibration certificate.
Table 19 HMP115/T probe measurement performance
Property Description/Value
Relative humidity
Measurement range 0 … 100 %RH
Accuracy in temperature range 0 … +40 °C (+32 … +104 °F)
1)
0 … 90 %RH
90 … 100 %RH
±1.5 %RH
±2.5 %RH
Accuracy in temperature range −40 … 0 °C, +40 … +60 °C (−40 … +32 °F, +104 … +140 °F)
1)
0 … 90 %RH ±3.0 %RH
90 … 100 %RH ±4.0 %RH
Factory calibration uncertainty at +20 °C (68 °F)
2)
0 … 40 %RH
40 … 75 %RH
±0.6 %RH
±1.0 %RH
Humidity sensor HUMICAPâ 180R
Stability ±2 %RH over 2 years
Temperature
Measurement range −40 … +60 °C (−40 °F … +140 °F)
Accuracy over temperature range
at 0 … +40 °C (+32 … +104 °F)
at −40 … 0 °C, +40 … +60 °C (−40 … +32 °F,
+104 … +140 °F)
± 0.2 °C (0.36 °F)
± 0.4 °C (0.72 °F)
Factory calibration uncertainty
2)
± 0.1 °C (0.18 °F)
RFL100 User Guide M211861EN-D
64