5.4.6 Adjusting measurement using Insight software
Figure 21 Adjusting relative humidity measurement in Insight
Computer with Windows operating system and Vaisala Insight software
installed
Vaisala USB cable 219690 for connecting the probe
Reference environments for the desired calibration and adjustment points (see
Adjustment points and requirements (page 44))
This procedure can be used to adjust the humidity or temperature measurement of the probe.
If you want to adjust both, adjust temperature measurement first, then repeat the procedure
for humidity.
1. Disconnect the probe from the data logger.
2. Connect the probe to Insight. See Connecting probe to Insight software (page 37).
3. Select
> Calibrate > Yes to switch the probe to calibration mode.
In calibration mode, the device will not use functions that may interfere with calibration
and adjustment.
4. Select the type of adjustment to perform: RH adjustment or T adjustment.