9. Select [3] Point 1 > Set. The meter now shows the currently measured value of the
selected parameter. Set the reference value using the arrow buttons and select OK.
Correction to measurement at point 1 is now shown in the text for menu item [3] Point 1. If
you are only doing a 1-point calibration, skip to step 12.
10. Place the probe in the second reference environment (second calibration point). Wait 20 –
40 minutes for the reading to stabilize.
Chapter 5 – Maintenance
47