Manual for VAKI Micro/Macro
Page 18
Information: Technical information about the
counter.
Figure 16 - Information
In the above information field, it is important to note the following information:
Light strength shows the brightness measured by the camera. The value should typically
be 1000-1400. If the value comes below 700, it must be checked if it can be improved by
cleaning the mirrors and glass. Another option may be to replace the bulb, or a VAKI service
technician may need to be called in to adjust the camera. Low light strength will also be
displayed if the bulb has not reached it maximum operating temperature. After washing or
other measures, the counter must be calibrated to obtain new brightness value.
Scan rate shows how many scans the camera performs each second. This value should be
~ 1116
Example of load indicator:
Figure 17 - Load indicator