Method Setup
03-914642-00:6
5-3
Parameter Range Default Description
Stir YES/NO NO
This parameter turns the Soil mechanical
stirrer on during purge time, or the water
mechanical stirrer on for a sample mixing
prior to sampling.
W. Stir Time
(min.)
0-9.9
minutes
0.0 min
The Stir motor will run for the length of
time entered for this parameter.
Note: This is a water method only
setting.
W. Settle
Time (sec)
0-99 seconds 00sec
This time allows particulates in the
sample to settle back to the bottom of the
vial after the Stir process. This time is
used after stirring a Water sample.
Syringe Flush 0 -20 YES
This parameter allows hot water to be
flushed in the 26 mL syringe and out to
waste.
PreHeat YES / NO YES
This parameter will allow the sample to
heat prior to purging. The heating will
occur after the water is added. Note: This
is a soil method only setting.
PreHeat Temp
amb/140
C40
C
This parameter sets the temp for the
preheating of the sample. Note: This is a
soil method only setting.
PreHeat Time 0-99.9 min. 3-0 min.
This parameter sets the time for
preheating of the sample. Note: This is a
soil method only setting.
Purge Time
(min.)
0-999.9
minutes
11.0 min
This is the time the sample will be purged.
This parameter must match the time
programmed on the purge and trap
concentrator. Note: This is a soil
method only setting.
Desorb Time
(min)
0-99.9
minutes
2.0 min
This parameter must match the time
programmed on the purge and trap
concentrator. The purge vessel is auto
drained back to the Archon waste bottle
during desorb.