Method Setup
03-914642-00:6
5-9
Sequence Step Purpose
21. Syringe -> waste The syringe now transfers the water to the waste bottle.
22. Helium ->waste Helium is flushed through the sample path to the waste
bottle.
23. Flush complete. Syringe flushing is complete.
24. Return vial The Robotic Arm verifies the vial position is empty, then
picks up the vial and returns it to the sample tray.
25. Parking The Robotic Arm goes to rear park.
26. Backflush water probe The water probe will now be flushed with hot water.
27. Elevator up The elevator raises the sampling stations.
28. Heating water Water is heated.
29. Water->water probe The backflush with hot water now begins.
30. Helium ->water probe After the water backflush is complete the probe is now
flushed with helium.
31. Elevator down The elevator moves down.
32. Autodrain The Archon waste valve opens and helium pressure
from the purge and trap will now drain the sparge tube.
33. Drain sparge tube The Archon will initiate a safety drain.
34. Sparge tube -> syringe If any sample water remains in the sparge tube, it is
drawn into the syringe.
35. Syringe -> waste The syringe transfers the water to the waste bottle.
36. Rinse sparge tube The sparge tube is rinsed if rinses have been
programmed.
37. Heating water Water is heated.
38. Water -> lines Hot water fills the lines.
39. Water -> syringe Hot water fills the syringe.
40. Syringe -> sparge tube The syringe transfers the water to the sparge tube.
41. Sparge tube -> syringe The syringe withdraws the water from the sparge tube.
42. Syringe -> waste The syringe transfers the water to the waste bottle.