9. Product Specifications
56 EzRay Air Wall (VEX-S300W) User Manual
To assure line voltage quality, a separate 3-core
grounded power cable connected directly to a central
distribution panel with an over-current circuit breaker
rated for 10A must be used.
The mains resistance should not exceed 0.5 Ω.
The system is available with a fixed tube current specification
based on the user's choice.
Power Supply is specified as a part of the ME EQUIPMENT.
Power plugs may have various specifications for each
country.
9.4 Environmental Specifications
Item Description
During operating
Temperature
10 ~ 35 ℃
Relative humidity 30 ~ 75 %
Atmospheric pressure 860 ~ 1060 hPa
Transport and storage
Temperature
-10 ~ 60 ℃
Relative humidity 10 ~ 75 % non-condensing
Atmospheric pressure 860 ~ 1060 hPa