5. Troubleshooting
EzRay Air Wall (VEX-S300W) User Manual 39
ENGLISH
5. Troubleshooting
In instances of abnormal operation, error messages will be displayed on the Control
Panel. If a problem persists, please request assistance from the customer support
information services.
Error Messages
A.0X: A problem occurred, and the system performs the
correction automatically. This alarm clears after the correction
is completed.
E.0X: An error occurred. Turn the power off, and then turn it
back on. If the error persists, contact your Service
Representative.
Error Code Check Parameter Description
E.02
X-ray Generator
An error related to X-ray exposure is not
possible to exposure X-ray in the state while
"E.02", "E.03", "E.04", "E.05" where power is
maintained.
After an X-ray exposure-related error occurs,
when the equipment is turned off and turned on,
X-ray exposure is normally performed.
E.03
E.04
E.05
A.06 X-ray Generator
Appears when the system needs cooling time
due to continuous operation. This alarm clears
when the system temperature goes down to
normal.
A.07 System
Appears when the Remote Exposure Switch has
been held down. Release the switch.
A.08 User
Appears when the exposure button has been
pressed and released before the X-ray exposure
finished with the default exposure time. Press
and hold the exposure button for the duration of
the exposure time.