8. Acquiring CEPH Images (Optional)
Green Smart User Manual 93
ENGLISH
5. Use the Column UP/DOWN button or switch option to adjust the height of the CEPH unit
to approximately match the height of the patient.
After adjusting the height of the column, align the Ear Rods to the
patient.
6. During the operation, properly align the Ear Rods to the patient's ears, so his/her head
does not move.
7. Align horizontally, so the patient's Frankfurt Line is parallel to the floor.
8. Direct the patient to swallow first before closing his/her mouth and to remain in his/her
current position until image acquisition is completed.
9. Click the READY button. The x-ray exposure has not started yet.
10. Now go to 8.4 X-ray Exposure to start the exposure.