9. Acquiring CBCT Images
112 Green Smart User Manual
Normal Patient Positioning
1. Insert the Normal Bite into the Normal Chinrest and cover it with a Sanitary Vinyl Cover.
The Sanitary Vinyl Cover is for single use only. It should be replaced
for each patient. Be sure to use the approved vinyl cover.
Clean the Chinrest and the Bite with ethanol and wipe with a dry
towel before the next patient.
2. Use the Temple Supports OPEN/CLOSE button on the control panel to widen the Temple
Supports.
3. Guide the patient to the equipment.
4. Use the Column UP/DOWN button or switch option to adjust the height of the equipment
so that the patient's chin reaches the Chinrest.
5. Guide the patient to stand in the center of the equipment and direct them to remain in the
position outlined below.
Hold the handles tightly.
Press the chest against the equipment.
Keep both feet close to the inside of the base.
Keep both shoulders parallel.
Straighten the Cervical Spine and stand still.
6. Let the patient bite the Bite along its grooves with his/her front teeth.