15. Appendices
158 Green Smart User Manual
DAP (Dose Area Product) Calculation
When you need more information on DAP measurement procedures or
test results for the equipment, please contact VATECH service center
or your local VATECH representative and get assistance from
VATEC H-authorized technicians
Measurement Overview
Results
Mode Exposure Condition
DAP
[mGy·cm
2
]
PANO Adult Man Normal (High
Resolution)
74 kVp / 12.0 mA / 13.5 s 110
PANO Child Normal (High
Resolution)
67 kVp / 10.0 mA / 11.5 s 55
PANO Adult Man Normal (Green) 74 kVp / 14.0 mA / 7.0 s 65
PANO Child Normal (Green) 67 kVp / 12.0 mA / 6.7 s 35
PANO Adult Man Insight PAN 67 kVp / 7.0 mA / 11.5 s 403
PANO Child Insight PAN 90 kVp / 15.0 mA / 3.9 s 295
CEPH Adult Man LAT (High
Resolution)
90 kVp / 15.0 mA / 3.9 s 41
CEPH Child LAT (High Resolution) 86 kVp / 10.0 mA / 3.9 s 38
CEPH Adult Man LAT (Green) 90 kVp / 16.0 mA / 1.9 s 25
CEPH Child LAT (Green) 86 kVp / 16.0 mA / 1.9 s 23
CBCT FOV 10x8.5 Adult Man (High
Resolution)
94 kVp / 8.7 mA / 16.4 s 646
CBCT FOV 10x8.5 Adult Man
(Green)
79 kVp / 5.0 mA / 16.4 s 256