7. Acquiring PANO Images
78 Green Smart User Manual
8. Let the patient maintain the posture as follows:
Close the mouth.
Place the tongue on the roof of the mouth.
Close the eyes.
As shown in the picture, the support unit of the integrated Chinrest
should touch the patient's acanthion point.
Ask the patient to maintain his/her position until the operation is
completed.