EasyManua.ls Logo

Vatech PaX-500 - Appendix; Recommended X-Ray Exposure Table

Vatech PaX-500
66 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
VATECH Co., Ltd.
Release
V2.0.0
PaX-500
60
Appendix
1. Recommended X-Ray Exposure Table
1.1. Pano Standard
Gender/Figure Hard Normal Soft
Man 72 kVp, 8 mA 70 kVp, 8 mA 68 kVp, 8 mA
Woman 70 kVp, 8 mA 68 kVp, 8 mA 66 kVp, 8 mA
Child 66 kVp, 7 mA 66 kVp, 7 mA 66 kVp, 7 mA
1.2. Pano Special
Gender/Figure Hard Normal Soft
Man 72 kVp, 8 mA 70 kVp, 8 mA 68 kVp, 8 mA
Woman 70 kVp, 8 mA 68 kVp, 8 mA 66 kVp, 8 mA
Child 66 kVp, 7 mA 66 kVp, 7 mA 66 kVp, 7 mA
1.3. Normal Ceph
Gender/Figure Hard Normal Soft
Man 78 kVp, 9 mA 78 kVp, 9 mA 78 kVp, 9 mA
Woman 76 kVp, 9 mA 76 kVp, 9 mA 76 kVp, 9 mA
Child 74 kVp, 8 mA 74 kVp, 8 mA 74 kVp, 8 mA
1.4. Quick Ceph
Gender/Figure Hard Normal Soft
Man 85 kVp, 9 mA 85 kVp, 9 mA 85 kVp, 9 mA
Woman 83 kVp, 9 mA 83 kVp, 9 mA 83 kVp, 9 mA
Child 80 kVp, 9 mA 80 kVp, 9 mA 80 kVp, 9 mA