9. Installing Software
PaX-i Plus / Insight (Model: PCH-30CS) Installation Manual 117
English
Disabling the Packing Mode
Unless the packing mode is disabled, the equipment will not
operate even after it is turned on.
PaX-i Plus_Insight has a unique feature─ packing mode─
built into the system to prevent the unit from being damaged
while shipping and transporting. Thus, it is in the packing
mode by factory default. Disabling the packing mode at this
step is required for a successful installation.
1. Click the General tab in the control panel.
2. In the Networking option, select the Serial checkbox and set Serial port and
baud rate as follows and then click Manager.
If an error has occurred during connection, make sure the
COM port setting is correct as follows:
1) Run Device Manager.
2) Check the COM port assigned to virtual serial port under
com0com – serial port emulators.
3) If necessary, change the COM port in the Networking
option and click Save.
3. Enter the command PVER] to verify the current mode. Note that the equipment is now
in packing mode.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vatech PaX-i Insight and is the answer not in the manual?

Vatech PaX-i Insight Specifications

General IconGeneral
BrandVatech
ModelPaX-i Insight
CategoryMedical Equipment
LanguageEnglish

Related product manuals