79
11 Supplement
VEGAPULS 67 • Probus PA
36533-EN-170405
Approvals
Instruments with approvals can have dierent technical specications depending on the version.
For that reason the associated approval documents of these instruments have to be carefully
noted. They are part of the delivery or can be downloaded under www.vega.com, "Instrument
search (serial number)" as well as in the download area.
11.2 CommunicationProbusPA
Instrumentmasterle
The instrument master le (GSD) contains the characteristic data of the Probus PA instrument.
These data are, e.g. the permissible transmission rates as well as information on diagnostics values
and the format of the measured value outputted by the PA instrument.
A bitmap le is also provided for the Probus network planning tool. This le is installed automati-
cally when the GSD le is integrated. The bitmap le is used for symbolic indication of the PA instru-
ment in the conguration tool.
ID number
Each Probus instrument gets an unambiguous ident number (ID number) from the Probus user
organisation (PNO). This ID number is also included in the name of the GSD le. Optionally in ad-
dition to this manufacturer-specic GSD le, PNO also provides a general so-called prole-specic
GSD le. If the general GSD le is used, the sensor must be set to the prole-specic ident number
via the DTM software. By default, the sensor operates with the manufacturer-specic ID number.
When using the instruments on a segment coupler SK-2 or SK-3, no special GSD les are required.
The following table shows the instrument ID and the GSD le names for VEGAPULS radar sensors.
Device name Instrument ID GSDlename
VEGA Instrument class in
prole3.02
VEGA Prole-specic
VEGAPULS WL 61 0x0CDB 0x9702 PSWL0CDB.GSD PA139702.GSD
VEGAPULS 61 0x0BFC 0x9702 PS610BFC.GSD PA139702.GSD
VEGAPULS 62 0x0BFD 0x9702 PS620BFD.GSD PA139702.GSD
VEGAPULS 63 0x0BFE 0x9702 PS630BFE.GSD PA139702.GSD
VEGAPULS 65 0x0BFF 0x9702 PS650BFF.GSD PA139702.GSD
VEGAPULS 66 0x0C00 0x9702 PS660C00.GSD PA139702.GSD
VEGAPULS 67 0x0C01 0x9702 PS670C01.GSD PA139702.GSD
VEGAPULS SR 68 0x0CDC 0x9702 PSSR0CDC.GSD PA139702.GSD
VEGAPULS 68 0x0C02 0x9702 PS680C02.GSD PA139702.GSD
VEGAPULS 69 0x0BFA 0x9702 VE010BFA.GSD PA139702.GSD
Cyclicaldatatrac
The master class 1 (e.g. PLC) cyclically reads out measured values from the sensor during opera-
tion. The below block diagram below shows which data can be accessed by the PLC.