Digital multimeter
17 / 25
mode can continue to test, and at this time “RANGE”.“MAX/MIN” “HOLD” re-ge t the controlling
Caution:
a) When measuring resistance, all the powers within the tested circuits must be disconnected and the
capacitance should be sufficiently discharged.
b) When measuring the resistance, any occurrence of voltage may lead to inaccurate readings, and if the
250V protection voltage is exceeded, it may damage the meter or threaten the safety of the user.
c) When using the range of 600Ω, first to short-circuit the testleads and measure the resistance of the
lead wire and then deduct it in the actual measurement.
Open circuit voltage: 600mV.
Overload protection: DC 250V or AC peak value.