11.9 Finding heating tubes in the oor
For the best search results, the heating pipes should
be disconnected from the earth connection� Connect
the transmitter to the heater's metal pipe and an earth
connection�
Note:
Suitable earth connections are any power
outlet protective conductors�
Trace the pipe's course with the receiver
11.10 Locating a complete electrical installation
Switch off the fuses and power down the
whole system� This measurement requires
manipulating the main electrical distribution
system and must only be performed by an
electrician�
With the cable detector, all present outlets and cables
can be determined and checked for contact� Remove
the main distribution bridge between the protective con-
ductor “PE” and the neutral conductor “N”�
Connect the transmitter to the distribution main of “N”
and “PE”�
The neutral conductor can be traced through the com-
plete system with the receiver