www.vtiinstruments.com
EX1629 Command Set 167
vtex1629_get_lxi_limit_event_enabled
FUNCTION PROTOTYPE
ViStatus vtex1629_get_lxi_limit_event_enabled (ViSession vi, ViInt32 lxi, ViUInt32 enabled0, ViUInt32
enabled1, ViUInt32 enabled2);
FUNCTION PARAMETERS
vi = contains a session handle to the instrument. This handle is obtained by the function and remains valid until the
session is closed.
lxi = the number of the LXI Trigger Bus line to configure. Valid input values: 0 to 7.
enabled0 = the limit mask for the channels 0 through 15, where the lower 16 bits are minimums for these channels,
while the upper 16 bits are maximums.
enabled1 = the limit mask for the channels 16 through 31, where the lower 16 bits are minimums for these channels,
while the upper 16 bits are maximums.
enabled2 = the limit mask for the channels 32 through 47, where the lower 16 bits are minimums for these channels,
while the upper 16 bits are maximums.
DATA ITEM RESET VALUE
Not applicable to this function.
DESCRIPTION
This function returns which measurement channels are enabled for Limit Events on the specified LXI Trigger Bus
line. Implemented in firmware version 1.4.0.